Document Type
Article
Publication Date
10-5-2022
Abstract
Purpose: Through-time spiral GRAPPA is a real-time imaging technique that enables ungated, free-breathing evaluation of the left ventricle. However, it requires a separate fully-sampled calibration scan to calculate GRAPPA weights. A self-calibrated through-time spiral GRAPPA method is proposed that uses a specially designed spiral trajectory with interleaved arm ordering such that consecutive undersampled frames can be merged to form calibration data, eliminating the separate fully-sampled acquisition. Theory and Methods: The proposed method considers the time needed to acquire data at all points in a GRAPPA calibration kernel when using interleaved arm ordering. Using this metric, simulations were performed to design a spiral trajectory for self-calibrated GRAPPA. Data were acquired in healthy volunteers using the proposed method and a comparison electrocardiogram-gated and breath-held cine scan. Left ventricular functional values and image quality are compared. Results: A 12-arm spiral trajectory was designed with a temporal resolution of 32.72 ms/cardiac phase with an acceleration factor of 3. Functional values calculated using the proposed method and the gold-standard method were not statistically significantly different (paired t-test, p < 0.05). Image quality ratings were lower for the proposed method, with statistically significantly different ratings (Wilcoxon signed rank test, p < 0.05) for two of five image quality aspects rated (level of artifact, blood-myocardium contrast). Conclusions: A self-calibrated through-time spiral GRAPPA reconstruction can enable ungated, free-breathing evaluation of the left ventricle in 71 s. Functional values are equivalent to a gold-standard cine technique, although some aspects of image quality may be inferior due to the real-time nature of the data collection.
Keywords
cardiovascular magnetic resonance, grappa, non-cartesian, parallel imaging, real-time, spiral
Language
English
Publication Title
Magnetic Resonance in Medicine
Grant
1563805
Rights
© 2022 The Authors. This is an open access work distributed under the terms of the Creative Commons Attribution-Non-Commercial (https://creativecommons.org/licenses/by-nc/4.0/) which permits non-commercial reuse, distribution, and reproduction in any medium, provided the original work is properly cited.
Creative Commons License

This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License
Recommended Citation
Franson D, Ahad J, Liu Y, et al. Self-calibrated through-time spiral GRAPPA for real-time, free-breathing evaluation of left ventricular function. Magn Reson Med. 2023; 89: 536-549. doi:10.1002/mrm.29462
Manuscript Version
Final Publisher Version