Document Type
Article
Publication Date
4-15-2022
Abstract
Purpose: This work proposes principal component analysis (PCA) coil compression and weight sharing to reduce acquisition and reconstruction time of through-time radial GRAPPA. Methods: Through-time radial GRAPPA enables ungated free-breathing motion-resolved cardiac imaging but requires a long calibration acquisition and GRAPPA weight calculation time. PCA coil compression reduces calibration data requirements and associated acquisition time, and weight sharing reduces the number of unique GRAPPA weight sets and associated weight computation time. In vivo cardiac data reconstructed with coil compression and weight sharing are compared to a gold standard to demonstrate improvement in calibration acquisition and reconstruction performance with minimal loss of image quality. Results: Coil compression from 30 physical to 12 virtual coils (90% of signal variance) decreases requisite calibration data by 60%, reducing calibration acquisition time to 6.7 s/slice from 31.5 s/slice reported in original through-time radial GRAPPA work. Resulting images have small increase in RMS error (RMSE). Reconstruction with a weight sharing factor of 8 results in eight-fold reduction in GRAPPA weight calculation time with a comparable RMSE to reconstructions with no weight sharing. Optimized parameters for coil compression and weight sharing applied to reconstructions enables images to be collected with a temporal resolution of 66 ms/frame and spatial resolution of 2.34 × 2.34 mm while reducing calibration acquisition time from 34 to 6.7 s, weight calculation time from 200 to 3 s, and weight application time 18 to 5 s. Conclusion: Coil compression and weight sharing applied to through-time radial GRAPPA enables fast free-breathing ungated cardiac cine without compromising image quality.
Keywords
cardiac imaging, channel compression, grappa, non-cartesian, parallel imaging
Language
English
Publication Title
Magnetic Resonance in Medicine
Grant
1563805
Rights
© 2022 The Authors. This is an open access work distributed under the terms of the Creative Commons Attribution-Non-Commercial (https://creativecommons.org/licenses/by-nc/4.0/) which permits non-commercial reuse, distribution, and reproduction in any medium, provided the original work is properly cited.
Creative Commons License

This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License
Recommended Citation
Ahad J, Cummings E, Franson D, Hamilton J, Seiberlich N. Optimization of through-time radial GRAPPA with coil compression and weight sharing. Magn Reson Med. 2022; 88(3): 1244-1254. doi:10.1002/mrm.29258
Manuscript Version
Final Publisher Version